Description
The BBS 1150 blackbody calibration source features a
high emissivity of +0.995 with a large 2” aperture and
superior uniformly heated spherical cavity.
Simply adjust the set point on the precision digital PID
controller to any temperature from 200°C to 1150°C
(392°F to 2102°F).
The Model BBS 1150 will attain the set point
temperature stabilize and produce the radiometric
temperature readings necessary for precision calibration
of non-contact infrared thermometers, thermal imaging
detectors/camera, focal plane array detectors, heat flux
meters, spectrophotometers and optical devices.
Two Blackbodies perfect for mid to high temperature applications. User selectable temperature set points within 200° to 1150°C (BBS 1150) or 300° to 1700°C (BBS 1700). Simply adjust the set point on the precision digital PID controller to any temperature within the specified temperature ranges and the two BlackBodies should stablize and begin reading.
Process Sensors blackbody calibration source Model BBS 1700 features a high emissivity of +0.99 and a 1″ (25mm) aperture. The PSC-BBS 1700 delivers a temperature ranging from 300°C to 1700°C. To change the temperature, you adjust the set point on the precision digital PID controller to any temperature. But, within the specified temperature range. Upon reaching the user selectable set point temperature, the model BBS 1700 will stabilize and produce a temperature reading to within 0.5°C. This assures high accuracy calibration of non-contact infrared thermometers, thermal imaging camera systems, focal plane array detectors, etc.
The PSC-BBS 1700 brochure explains more technical data as well.
Temperature range- 300° to 1700°C or 572° to 3092°F
Accuracy- ±0.25% of reading ±1°C
Temperature Resolution- 1 degree
Stability- 1°C per 8 hour period
Emitter Diameter- 25mm (1.0”)
Emissivity (effective)- +0.99
Sensor- Precision platinum thermocouple
Method of Control- Digital self-tuning PID controller
Warm-up Time- 1 hour from room temperature to 1000°C
Operating Ambient Temperature- 0° to 44°C (32° to 110°F)
Cooling- Fan cooled, air inlet on rear panel
Power Requirements- 208VAC ±10% 50/60Hz or 3.0kw max. (230VAC optional)
Dimensions (WxHxD)- 19” x 25” x 21”
Weight- 80kg (175 lbs.)
Optional- 115/230 VAC, 3 kwatts max.
Temperature Range- 200° to 1150°C or 392° to 2102°F
Accuracy- ±0.25% of reading ±1°C
Temperature Resolution- 1 degree
Stability- 0.5°C
Emitter Diameter- 2.0” (51mm)
Emissivity (Effective)- >0.99
Method of Measurement- Precision thermocouple
Method of Control- Precision PID controller
Warm-up Time- 1 hour from room temperature to 1000°C
Operating Ambient Temperature- 0° to 45°C (32° to 113°F)
Cooling- Fan cooled, air exhaust at rear panel
Power Requirements- 230VAC 50/60Hz, 4.46kw, 19 amps maximum.
Dimensions (WxHxD)- 19” x 25” x 21” (482.6 x 635 x 533.4mm)
Weight-80kg (175 lbs.)
The PSC-BBS 1150 brochure explains more technical data as well.